ModelFI-C40F
TypePipe type
AttachmentFI-CF1/3, FI-CF2/4
Detection principleRefractive index (surface light source)
Rated concentration rangeBrix: 0–40% (nD: 1.32500–1.41000)
Display methodStatus indicator light
Display rangeBrix: 0–50%
Supported fluidsNon-corrosive aqueous solutions (such as water-soluble coolants, mold release agents, etc.)*1
Supported fluid temperature0–70°C 32–158°F (no freezing)
Connection diameterFI-CF1: Rc3/4 (20 A); FI-CF3: NPT3/4FI-CF2: Rc1 1/2 (40 A); FI-CF4: NPT1 1/2
Rated pressure range1.0 MPa or less
Pressure resistance2.0 MPa
Display resolutionBrix: 0.01/0.1% (default value: 0.1)(nD: 0.00001)
Response time1.0 s / 2.5 s / 5.0 s / 10.0 s / 30.0 s / 60.0 s / 120.0 s / 200.0 s
Measurement accuracyBrix: ±0.2%*2*3 (nD: ±0.0003)
Concentration unitBrix nD*4
Temperature measurement accuracy±1.0°C 1.8°F*2
Power supplyCurrent consumption25 mA or less
Environmental resistanceEnclosure ratingIP65/IP67 (IEC60529)
Ambient temperature–10 to +60°C 14 to +140°F (no freezing)
Relative humidity35–85% RH (no condensation)
Vibration resistance10–500 Hz; Power spectral density: 0.816 G2/Hz; X, Y and Z directions
Shock resistance100 m/s2 (approx. 10 G), 16 ms pulses, 1000 times each for X, Y and Z directions
MaterialLiquid end materialsSensor element: sapphireBody: SCS16A; Gasket: FFKM
Other materialsBody: SUS304; PPS indicator light: PPSU / TPU / PBTCable: PVC
WeightApprox. 410 g 14.46 oz
*1 Use water as the solvent, and use materials that are soluble in water. If the particles are not water soluble, such as with slurry, the refractive index may not change.*2 This is the guaranteed value from verification performed at KEYENCE inspection facilities. Measurement error may occur depending on the type, condition and temperature of the fluid used by the customer, as well as other factors.*3 This is the value obtained when sucrose solution is used in a constant 20°C 68°F environment, and absolute error and repeatability is taken into account.*4 When using solutions other than sucrose solutions, concentration can be adjusted via span adjustment.

Thông tin bổ sung

Thương hiệu
Thời gian bảo hành

1 Năm

Series
Yêu cầu tư vấn0983 995 440(Call/Zalo)

Có thể bạn thích…

Cam kết và hỗ trợ

Luôn có giá tốt nhất thị trường
Giao hàng trên toàn quốc
Hỗ trợ dịch vụ kỹ thuật

Khu vực Miền Nam

Call: 0386 334 447 / Chat Zalo zalo

Khu vực Miền Bắc

Call: 0384 988 968 / Chat Zalo zalo

Khu vực Miền Trung

Call: 0374 687 899 / Chat Zalo zalo

Dịch vụ kỹ thuật/Bảo hành

Call: 0348 957 778 / Chat Zalo zalo

Email liên hệ

Dịch vụ cung ứng sản phẩm

Vì sản phẩm đa dạng, nếu Quý khách chưa tìm được. Vui lòng liên hệ để được hỗ trợ.

Tư vấn lựa chọn hoặc chuyển đổi
Nhập khẩu thiết bị theo yêu cầu

dich-vu-sua-chua-bien-tan-thiet-bi-tu-dong-hoa