Product characteristics |
---|
Number of inputs and outputs | Number of digital outputs: 4 |
Factory setting | 2 switching outputs level (OUT1, OUT2); 2 switching outputs temperature (OUT3, OUT4); mineral oils |
Probe length L [mm] | 472 |
Application |
---|
Media | hydrous coolants; oils; oil-based media; water; media similar to water |
Dielectric constant of the medium | > 2 |
Cannot be used for | extremely conductive media; adhering media; granulates; bulk material; acids; alkali |
Tank pressure [bar] | 0.5; (when mounting with mounting accessories: E43001 – E43007, E43019) |
Coolants |
---|
Medium temperature [°C] | 0…35; (with climatic tube E43101 35…60 °C) |
Oil |
---|
Medium temperature [°C] | 0…70 |
Medium temperature short time [°C] | 0…90; (< 1 h) |
Water |
---|
Medium temperature [°C] | 0…35; (with climatic tube E43101 35…60 °C) |
Electrical data |
---|
Operating voltage [V] | 18…30 DC |
Current consumption [mA] | < 50 |
Protection class | III |
Reverse polarity protection | yes |
Power-on delay time [s] | < 3 |
Inputs / outputs |
---|
Number of inputs and outputs | Number of digital outputs: 4 |
Outputs |
---|
Total number of outputs | 4 |
Output signal | switching signal; IO-Link |
Electrical design | PNP/NPN |
Number of digital outputs | 4 |
Output function | normally open / normally closed; (parameterisable) |
Max. voltage drop switching output DC [V] | 2.5 |
Permanent current rating of switching output DC [mA] | 200 |
Short-circuit protection | yes |
Type of short-circuit protection | thermal, pulsed |
Overload protection | yes |
Measuring/setting range |
---|
Factory setting | 2 switching outputs level (OUT1, OUT2); 2 switching outputs temperature (OUT3, OUT4); mineral oils |
Probe length L [mm] | 472 |
Active range A [mm] | 390 |
Inactive range I1 / I2 [mm] | 60 / 22 |
Setting range |
---|
Set point SP [mm] | 35…390 |
Reset point rP [mm] | 30…385 |
In steps of [mm] | 5 |
Reference point OP [mm] | 139 – 163 – 188 – 212 – 236 – 261 – 285 – 310 – 334 – 358 – 383 – 407 – OFF |
Hysteresis, OP [mm] | 3 |
Temperature monitoring |
---|
Measuring range [°C] | 0…90 |
Set point SP [°C] | 0.5…90 |
Reset point rP [°C] | 0…89.5 |
In steps of [°C] | 0.5 |
Resolution of switching output [K] | 0.5 |
Accuracy / deviations |
---|
Measuring error [% of the final value] | ± 5 |
Repeatability | ± 2 |
Resolution [mm] | 5 |
Temperature monitoring |
---|
Switch point accuracy [K] | ± 1,5 |
Resolution [K] | 0.5 |
Dynamic response T09 [s] | 90 |
Software / programming |
---|
Parameter setting options | hysteresis / window; normally open / normally closed; SP/rP position; OP position; adjustment OP; medium selection; offset setting; Display unit; process value assignment |
Interfaces |
---|
Communication interface | IO-Link |
Transmission type | COM2 (38,4 kBaud) |
IO-Link revision | 1.1 |
SDCI standard | IEC 61131-9 |
Profiles | Smart Sensor: Process Data Variable; Device Identification, Device Diagnosis |
SIO mode | yes |
Required master port type | A |
Process data analogue | 4 |
Process data binary | 4 |
Min. process cycle time [ms] | 4.5 |
Supported DeviceIDs | Type of operation | DeviceID | Default | 19 |
|
Operating conditions |
---|
Ambient temperature [°C] | 0…60 |
Storage temperature [°C] | -25…80 |
Protection | IP 67 |
Mechanical data |
---|
Weight [g] | 443.5 |
Dimensions [mm] | Ø 16 |
Materials | stainless steel (1.4301 / 304); stainless steel (1.4404 / 316L); FKM; NBR; PBT; PC; PP; PA; TPV |
Materials (wetted parts) | PP |
Displays / operating elements |
---|
Display | Display unit | 4 x LED, green (cm, inch, °C, °F) | switching status | 4 x LED, yellow (OUT4…OUT1) | measured values | alphanumeric display, 4-digit | parameter setting | alphanumeric display, 4-digit |
|
Remarks |
---|
Notes | For 8-pole sockets the core colours are not standardised.; Please note the wiring of the sensor and the sockets (see data sheet).; Please see the technical note under “Downloads” |
Pack quantity | 1 pcs. |
Electrical connection |
---|
Connection | Connector: 1 x M12 |