|  Product characteristics | 
|---|
|  Number of inputs and outputs |  Number of digital outputs: 2; Number of analogue outputs: 1 | 
|  Process connection |  diameter Ø 8 mm | 
|  Application | 
|---|
|  Special feature |  Gold-plated contacts | 
|  Media |  water; glycol solutions; air; oils | 
|  Note on media | |  low-viscosity oils with viscosity: ≤ 40 mm²/s (104 °F) |  |  high-viscosity oils with viscosity: > 40 mm²/s (104 °F) |  
  | 
|  Medium temperature [°F] |  -4…212 | 
|  Pressure rating [bar] |  50 | 
|  Pressure rating [psi] |  725 | 
|  MAWP (for applications according to CRN) [bar] |  50 | 
|  Electrical data | 
|---|
|  Operating voltage [V] |  18…30 DC | 
|  Current consumption [mA] |  < 100 | 
|  Protection class |  III | 
|  Reverse polarity protection |  yes | 
|  Power-on delay time [s] |  10 | 
|  Inputs / outputs | 
|---|
|  Number of inputs and outputs |  Number of digital outputs: 2; Number of analogue outputs: 1 | 
|  Outputs | 
|---|
|  Total number of outputs |  2 | 
|  Output signal |  switching signal; analogue signal; frequency signal; IO-Link; (configurable) | 
|  Electrical design |  PNP/NPN | 
|  Number of digital outputs |  2 | 
|  Output function |  normally open / normally closed; (parameterisable) | 
|  Max. voltage drop switching output DC [V] |  2.5 | 
|  Permanent current rating of switching output DC [mA] |  250 | 
|  Number of analogue outputs |  1 | 
|  Analogue current output [mA] |  4…20; (scalable) | 
|  Max. load [Ω] |  350 | 
|  Short-circuit protection |  yes | 
|  Type of short-circuit protection |  pulsed | 
|  Overload protection |  yes | 
|  Frequency of the output [Hz] |  0…1000 | 
|  Measuring/setting range | 
|---|
|  Probe length L [mm] |  200 | 
|  Operating mode |  relative; absolutely liquid; absolutely gaseous; (absolute: reference measurement recommended) | 
|  Note on factory setting |  | 
|  Liquids | 
|---|
|  Setting range [ft/s] |  0…19.5 | 
|  Greatest sensitivity [ft/s] |  0.15…9.85 | 
|  Gases | 
|---|
|  Setting range [ft/s] |  0…656 | 
|  Greatest sensitivity [ft/s] |  6…328 | 
|  Temperature monitoring | 
|---|
|  Measuring range [°F] |  -4…212 | 
|  Resolution [°F] |  0.5 | 
|  Accuracy / deviations | 
|---|
|  Flow monitoring | 
|---|
|  Temperature drift [cm/s x 1/K] |  0,01 fps x 1/K (< 68 °F; > 158 °F) | 
|  Temperature gradient [K/min] |  100 | 
|  Accuracy |  ± (7 % MW + 2 % MEW); (for relative mode in the range of maximum sensitivity under the following conditions :; water: 68…158 °F; inlet length: 5 ft; DN25 (DIN 2448); mounting position according to instructions; Accuracy can differ for other media and mounting positions.) | 
|  Repeatability |  0,05 m/s; (water; flow velocity: 0,05…3 m/s) | 
|  Temperature monitoring | 
|---|
|  Temperature drift |  ± 0,003 K/°F | 
|  Accuracy [K] |  ± 0,3 / ± 1; (water; flow velocity: 1…9,85 fps / air; flow velocity: > 32,8 fps) | 
|  Response times | 
|---|
|  Flow monitoring | 
|---|
|  Response time [s] |  0.5; (T09; water; glycol: 0,8 s; air: 7 s; oil: 1,8 s; each T09) | 
|  Temperature monitoring | 
|---|
|  Dynamic response T05 / T09 [s] |  1,5 (T09); (water; flow velocity: 1…9,85 fps) | 
|  Software / programming | 
|---|
|  Parameter setting options |  hysteresis / window; normally open / normally closed; switching logic; current/frequency output; medium selection; Damping; Teach function; display can be rotated and switched off; standard unit of measurement; process value colour | 
|  Interfaces | 
|---|
|  Communication interface |  IO-Link | 
|  Transmission type |  COM2 (38,4 kBaud) | 
|  IO-Link revision |  1.1 | 
|  SDCI standard |  IEC 61131-9 | 
|  Profiles |  Smart Sensor: Process Data Variable; Device Identification, Device Diagnosis | 
|  SIO mode |  yes | 
|  Required master port type |  A | 
|  Process data analogue |  2 | 
|  Process data binary |  2 | 
|  Min. process cycle time [ms] |  3 | 
|  Supported DeviceIDs | |  Type of operation |  DeviceID |  |  Factory setting / ModE = (REL) |  537 |  |  ModE = (LIQU) |  544 |  |  ModE = (GAS) |  551 |  
  | 
|  Operating conditions | 
|---|
|  Ambient temperature [°F] |  -40…176 | 
|  Storage temperature [°F] |  -40…212 | 
|  Protection |  IP 65; IP 67 | 
|  Mechanical data | 
|---|
|  Weight [g] |  344.5 | 
|  Materials |  stainless steel (1.4404 / 316L); PBT-GF20; PBT-GF30 | 
|  Materials (wetted parts) |  stainless steel (1.4404 / 316L) | 
|  Process connection |  diameter Ø 8 mm | 
|  Displays / operating elements | 
|---|
|  Display | |  Display unit |  6 x LED, green (%, fps, gpm, cfm, °F, 10³) |  |  switching status |  2 x LED, yellow |  |  measured values |  alphanumeric display, red/green 4-digit |  
  | 
|  Remarks | 
|---|
|  Remarks | |  MW = measured value |  |  MEW = Final value of the measuring range |  
  | 
|  Pack quantity |  1 pcs. | 
|  Electrical connection | 
|---|
|  Connection |  Connector: 1 x M12; Contacts: gold-plated |