| Product characteristics |
|---|
| Number of inputs and outputs | Number of analogue outputs: 1 |
| Measuring range [°C] | -50…150 |
| Process connection | threaded connection G 1/2 |
| Installation length EL [mm] | 100 |
| Application |
|---|
| Special feature | Gold-plated contacts |
| Measuring element | 1 x Pt 1000; (to DIN EN 60751, class A) |
| Media | liquids and gases |
| Pressure rating [bar] | 160 |
| Note on pressure rating | | sensor | | When mounted in adapters the specifications of the adapter data sheet apply. |
|
| Electrical data |
|---|
| Operating voltage [V] | 18…32 DC; (“supply class 2” to cULus) |
| Current consumption [mA] | < 50 |
| Protection class | III |
| Reverse polarity protection | yes |
| Power-on delay time [s] | 1 |
| Inputs / outputs |
|---|
| Number of inputs and outputs | Number of analogue outputs: 1 |
| Outputs |
|---|
| Total number of outputs | 1 |
| Output signal | analogue signal; IO-Link; (configurable) |
| Number of analogue outputs | 1 |
| Analogue current output [mA] | 4…20 |
| Max. load [Ω] | 250; ((18…19 V); 19…32 V: 300 Ω) |
| Short-circuit protection | yes |
| Overload protection | yes |
| Measuring/setting range |
|---|
| Measuring range [°C] | -50…150 |
| Factory setting | 0…100 °C |
| Resolution |
|---|
| Resolution of analogue output [K] | 0.04 |
| Accuracy / deviations |
|---|
| Precision analogue output [K] | ± 0,3 + (± 0,1 % MS) |
| Temperature coefficient [% of the span / 10 K] | 0,1; (in case of deviation from the reference condition 25 ± 5 °C) |
| Response times |
|---|
| Dynamic response T05 / T09 [s] | 1 / 3 |
| Interfaces |
|---|
| Communication interface | IO-Link |
| Transmission type | COM2 (38,4 kBaud) |
| IO-Link revision | 1.1 |
| Operating conditions |
|---|
| Ambient temperature [°C] | -25…80 |
| Storage temperature [°C] | -40…100 |
| Protection | IP 67; IP 68; IP 69K |
| Mechanical data |
|---|
| Weight [g] | 131 |
| Materials | stainless steel (1.4404 / 316L); PEI; FKM |
| Materials (wetted parts) | stainless steel (1.4404 / 316L); O-ring: FKM |
| Process connection | threaded connection G 1/2 |
| Probe diameter [mm] | 6 |
| Installation length EL [mm] | 100 |
| Displays / operating elements |
|---|
| Display | | operating status | 1 x LED, green |
|
| Remarks |
|---|
| Remarks | |
| Pack quantity | 1 pcs. |
| Electrical connection |
|---|
| Connection | Connector: 1 x M12; Contacts: gold-plated |