| Product characteristics |
|---|
| Number of inputs and outputs | Number of digital outputs: 1; Number of analogue outputs: 1 |
| Measuring range | |
| Process connection | threaded connection G 1/2 external thread |
| Installation length EL [mm] | 87.5 |
| Application |
|---|
| Special feature | Gold-plated contacts |
| Measuring element | 1 x Pt 1000 + 1 x NTC; (thermically coupled, with backup function (temperature measuring even if one element fails)) |
| Media | liquids and gases |
| Pressure rating [bar] | 50 |
| Minimum installation depth [mm] | 25 |
| MAWP (for applications according to CRN) [bar] | 50 |
| Electrical data |
|---|
| Operating voltage [V] | 18…32 DC; (“supply class 2” to cULus) |
| Current consumption [mA] | 6; (24 V) |
| Protection class | III |
| Reverse polarity protection | yes |
| Power-on delay time [s] | 8 |
| Integrated watchdog | yes |
| Inputs / outputs |
|---|
| Number of inputs and outputs | Number of digital outputs: 1; Number of analogue outputs: 1 |
| Outputs |
|---|
| Total number of outputs | 2 |
| Output signal | switching signal; analogue signal; IO-Link; (configurable) |
| Electrical design | PNP/NPN |
| Number of digital outputs | 1 |
| Output function | normally open / normally closed / diagnostic signal; (parameterisable) |
| Max. voltage drop switching output DC [V] | 2 |
| Permanent current rating of switching output DC [mA] | 150; (200 (…60°C); 250 (…40 °C)) |
| Diagnostic output | diagnostic output for drift and fault monitoring |
| Number of analogue outputs | 1 |
| Analogue current output [mA] | 4…20 |
| Max. load [Ω] | (Ub – 15 V) x 50 |
| Short-circuit protection | yes |
| Type of short-circuit protection | pulsed |
| Overload protection | yes |
| Measuring/setting range |
|---|
| Measuring range | |
| Note on measuring range | |
| Factory setting | 0…150 °C / 32…302 °F |
| Drift warning | |
| Drift alarm | |
| In steps of | |
| Resolution |
|---|
| Resolution of analogue output [K] | 0.05 |
| Accuracy / deviations |
|---|
| Precision analogue output [K] | ± 0,2 (-10…130°C); ± 0,3 (130…140°C); ± 0,3 ± 0,1% MS (-25…-10/140…160°C); (probe completely inserted into the measured medium up to the sealing chamfer) |
| Temperature coefficient [% of the span / 10 K] | < ± 0,01; (in case of deviation from the reference condition 25 ± 5 °C) |
| Response times |
|---|
| Dynamic response T05 / T09 [s] | 3 / 6 |
| Software / programming |
|---|
| Parameter setting options | drift warning/drift alarm threshold; Fail-Safe; Display unit; scaling of the analogue output; redundancy switching; performance diagnostic output; switching logic; normally open / normally closed |
| Interfaces |
|---|
| Communication interface | IO-Link |
| Transmission type | COM1 (4,8 kBaud) |
| IO-Link revision | 1.1 |
| SDCI standard | IEC 61131-9 CDV |
| Profiles | Smart Sensor: Process Data Variable; Device Identification, Device Diagnosis |
| SIO mode | yes |
| Required master port type | A |
| Process data analogue | 1 |
| Process data binary | 1 |
| Min. process cycle time [ms] | 18.8 |
| Supported DeviceIDs | | Type of operation | DeviceID | | Default | 323 |
|
| Operating conditions |
|---|
| Ambient temperature [°C] | -25…70 |
| Storage temperature [°C] | -40…85 |
| Protection | IP 68; IP 69K |
| Mechanical data |
|---|
| Weight [g] | 345 |
| Materials | stainless steel (1.4435 / 316L); stainless steel (1.4404 / 316L); PEI; FKM |
| Materials (wetted parts) | stainless steel (1.4435 / 316L); stainless steel (1.4404 / 316L) |
| Tightening torque [Nm] | 30…50 |
| Process connection | threaded connection G 1/2 external thread |
| Surface characteristics Ra/Rz of the wetted parts | Ra: < 0.6 |
| Installation length EL [mm] | 87.5 |
| Remarks |
|---|
| Remarks | |
| Pack quantity | 1 pcs. |
| Electrical connection |
|---|
| Connection | Connector: 1 x M12; Contacts: gold-plated |