|  Product characteristics | 
|---|
|  Number of inputs and outputs |  Number of digital outputs: 1; Number of analogue outputs: 1 | 
|  Measuring range |  | 
|  Process connection |  diameter Ø 6 mm | 
|  Installation length EL [mm] |  350 | 
|  Application | 
|---|
|  Special feature |  Gold-plated contacts | 
|  Measuring element |  1 x Pt 1000 | 
|  Reference element |  1 x NTC | 
|  Media |  liquids and gases | 
|  Pressure rating [bar] |  160 | 
|  Electrical data | 
|---|
|  Operating voltage [V] |  18…32 DC; (“supply class 2” to cULus) | 
|  Current consumption [mA] |  10; (24 V) | 
|  Protection class |  III | 
|  Reverse polarity protection |  yes | 
|  Power-on delay time [s] |  6 | 
|  Integrated watchdog |  yes | 
|  Inputs / outputs | 
|---|
|  Number of inputs and outputs |  Number of digital outputs: 1; Number of analogue outputs: 1 | 
|  Outputs | 
|---|
|  Total number of outputs |  2 | 
|  Output signal |  analogue signal; IO-Link; calibration check status | 
|  Electrical design |  PNP/NPN | 
|  Number of digital outputs |  1 | 
|  Output function |  normally open; (diagnostic signal) | 
|  Max. voltage drop switching output DC [V] |  2 | 
|  Permanent current rating of switching output DC [mA] |  100 | 
|  Diagnostic output |  calibration check status and error diagnostics | 
|  Number of analogue outputs |  1 | 
|  Analogue current output [mA] |  4…20 | 
|  Max. load [Ω] |  (Ub – 15 V) x 50 | 
|  Short-circuit protection |  yes | 
|  Type of short-circuit protection |  pulsed | 
|  Overload protection |  yes | 
|  Measuring/setting range | 
|---|
|  Measuring range |  | 
|  Note on measuring range |  | 
|  Factory setting |  -10…150 °C / 14…302 °F | 
|  Calibration check limit [K] |  0.5…3 | 
|  In steps of [K] |  0.05 | 
|  Resolution | 
|---|
|  Resolution of analogue output [K] |  0.05 | 
|  Accuracy / deviations | 
|---|
|  Precision analogue output [K] |  ± 0,2 | 
|  Precision IO-Link [K] |  ± 0,2 | 
|  Temperature coefficient analogue output [% of the span / 10 K] |  < ± 0,02; (in case of deviation from the reference condition 25 ± 5 °C) | 
|  Temperature coefficient IO-Link [% of the span / 10 K] |  < ± 0,01; (in case of deviation from the reference condition 25 ± 5 °C) | 
|  Response times | 
|---|
|  Dynamic response T05 / T09 [s] |  1,5 / 4 | 
|  Software / programming | 
|---|
|  Parameter setting options |  Display unit; scaling of the analogue output; calibration check limit; switching logic diagnostic output; simulation mode | 
|  Interfaces | 
|---|
|  Communication interface |  IO-Link | 
|  Transmission type |  COM2 (38,4 kBaud) | 
|  IO-Link revision |  1.1 | 
|  SDCI standard |  IEC 61131-9 CDV | 
|  Profiles |  Digital Measuring Sensor, Common Profile, Blob Transfer | 
|  SIO mode |  yes | 
|  Required master port type |  A | 
|  Process data analogue |  1 | 
|  Process data binary |  1 | 
|  Min. process cycle time [ms] |  4.4 | 
|  IO-Link resolution temperature [K] |  0.01 | 
|  Supported DeviceIDs | |  Type of operation |  DeviceID |  |  Default |  1129 |  
  | 
|  Operating conditions | 
|---|
|  Ambient temperature [°C] |  -25…70 | 
|  Note on ambient temperature | |  max. internal device temperature: 125 °C |  
  | 
|  Storage temperature [°C] |  -40…100 | 
|  Protection |  IP 68; IP 69K | 
|  Mechanical data | 
|---|
|  Weight [g] |  315.6 | 
|  Dimensions [mm] |  Ø 50 / L = 393.5 | 
|  Materials |  stainless steel (1.4404 / 316L); PEI; FKM; PFA | 
|  Materials (wetted parts) |  stainless steel (1.4404 / 316L); PEEK | 
|  Process connection |  diameter Ø 6 mm | 
|  Surface characteristics Ra/Rz of the wetted parts |  Ra: < 0.8 | 
|  Probe diameter [mm] |  6 | 
|  Installation length EL [mm] |  350 | 
|  Remarks | 
|---|
|  Remarks | |  MS = set measuring span |  |  operating voltage “supply class 2” according to cULus |  
  | 
|  Pack quantity |  1 pcs. | 
|  Electrical connection | 
|---|
|  Connection |  Connector: 1 x M12; Contacts: gold-plated |